Treatment Laser |
Type |
Diode pumped solid-state, mode-locked |
Center wavelength |
1030 (±5nm) |
Pulse energy |
1 - 10µJ |
Pulse duration |
<600fs |
Pulse rep. rate |
120kHz ±5%, with integer down sampling when modulated on |
CDRH laser class (21 CFR 1040) |
Class 4 |
Optical |
Numerical aperture |
-2 ≥ 0.125 (e ) |
Scan field, lateral, max |
≥14mm diameter, telecentric |
Video Imaging |
Type |
Monochrome CCD |
Frame rate |
>24 frames per second |
Field of view |
17mm +0/-1mm DIA |
Illumination |
735nm ±10 nm LED array, variable brightness |
Optical resolution |
<40µm |
OCT |
Type |
Spectral domain |
Center wavelength |
820 – 930nm, CW |
Resolution, lateral |
15µm or better at focus |
Resolution, axial |
30µm or better |
CDRH class (21 CFR 1040) |
Class 3R |
ANSI Z136.1 |
Class 1 |
Patient Interface – Aqueous Contact (LIQUID OPTICS Interface) |
Contact element |
Sterile buffered saline solution, such as Alcon BSS (Alcon P/N
351/55005-1) or equivalent (not supplied). |
Inner diameter |
12.0mm (LIQUID OPTICS Interface-12) |
Fixation method |
Annular vacuum flange onto sclera |
Suction ring seal material |
Medical grade silicone elastomer |
Vacuum pressure |
300 – 700mmHg (gauge) |
Electrical Requirements¡¡ |
Voltage |
200 – 240 VAC
50/60 Hz single-phase 15A |
Environmental Requirements (Operating)¡¡ |
Room condition |
59 – 89.6°F (15 – 32°C)
Up to 80% @ 89.6°F (32°C) non-condensing |
Operating Environment |
Temperature controlled environment with temperature controlled to
within ±3.6°F (±2°C) |
Environmental Requirements (Non-operating) |
Maximum altitude |
Standard commercial shipping altitude |
Temperature range |
14 – 131°F (-10 – 55°C) |
Maximum humidity |
Up to 80% @ 131°F (55°C) non-condensing |
Physical Characteristics¡¡ |
Height |
1.15m ±.025m |
Length |
1.64m ±.025m |
Width |
0.87m ±.025m |
Weight |
340kg (750lbs) |
Optical NOHD |
0m |
Laser Safety Eyewear |
Not required |
IO¡¡ |
Display |
GUI: touch-screen, flat panel |
Power cable length |
≥8.2 ft (2.5m), detachable |